Project Overview

These emails document communications between Maselli Misure S.p.A. and Sidel regarding measurement equipment for a project at San Giorgio in Bosco (referred to as "Diet As Regular"). The discussions focus on testing and configuring refractometers, ADAM modules, and conductivity equipment.

Key Equipment

  • UR62 refractometer with ADAM RS485/4-20mA converter
  • UR29 refractometer with ADAM RS485/4-20mA converter
  • Conductivity meter for water measurement

Testing Arrangements

  • Tests scheduled for May 6, 2025, at Maselli facilities
  • Testing may extend to a second day if needed
  • Equipment to be available for testing rather than shipped to Sidel

Sidel Responsibilities for Testing

  • PC with HMI machine interface
  • PLC machine
  • 24Vdc power supply
  • USB-RS485 converter (Moxa) for ADAM programming

Maselli Responsibilities for Testing

  • Instruments purchased for the project
  • Cables and required accessories

Technical Discussions

  • Significant focus on signal latency concerns
  • UR24 device in Tolosa reportedly provides measurements at 2-second intervals
  • Initial requirement of 300ms response time appears difficult to achieve
  • Tests show MP06 connected to two sensors cannot meet 300ms requirement even with increased serial speed (115200 baud vs. current 19200)
  • Discussions about using direct ADAM modules versus routing through MP06
  • Configuration options being considered:
    1. ADAM modules directly to PLC (preferred for minimizing latency)
    2. ADAM outputs to MP06, then to PLC via both Profibus and 4-20mA

Delivery Schedule

  • UR62, UR29 with ADAM module expected by April 28
  • Conductivity kit expected by May 6
  • Technical documentation for the equipment being shared

Documentation and Firmware Information

  • ADAM configuration procedure and URXX protocol format documentation
  • Module firmware with updates mentioned in "procedura v3"
  • Utility software "MAU_V_2_9_6.msi" provided for configuration
  • ADAM-4000 Series User Manual (Ed.9)
  • References to "Maselli_protocol_with_Adam.pdf" and "Maselli_protocol_with_Adam_2.pdf"
  • Firmware updates for ADAM modules included as part of the planned testing

Future Development and Speed Improvements

  • Discussions about developing a faster communication bus for future applications
  • Current limitation: MP06 connected to two sensors cannot meet 300ms requirement even at increased serial speed (115200 baud vs. standard 19200)
  • Problem identified as intrinsic to the master-slave protocol between MP06 and sensors
  • Historical context: Previous discussions with Paolo (retired software developer) indicated that the "poor speed" of UR24 would be resolved with new generation of instruments equipped with new microcontrollers
  • Impact analysis: For large machines (60,000 l/h), a 300ms latency equates to 5 liters of product flow at nominal speed
  • Current UR24 device in Tolosa reportedly provides measurements at 2-second intervals, causing significant uncertainty in syrup measurements for large machines
  • Matteo indicated new development was in progress but not yet ready (responded "Soon, soon" when asked for documentation on the new faster bus)
Tamaño máquina 6 9 10
Flujo nominal (l/h) 27000 54000 60000
Flujo mínimo (l/h) 13500 27000 30000
Equivalente 300ms a velocidad nominal (litros) 2,25 4,5 5
Equivalente 300ms a velocidad mínima (litros) 1,125 2,25 2,5

Índice de Mensajes


  • 33828f4d4ea4fe2b9c52f4ecfe240929

I: SIDEL (01011476) R: Nuova configurazione e banAndrea Maselli andreamaselli@maselli.comco di test latenza, offerta RDO250201 per UR29 con modulo Adam e conducibilità

  • 06-05-2025

Inoltro documento aggiornato (V2) Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Quarantelli

'Orsi, Walter [EXTERNAL]' Walter.Orsi.external@sidel.com; 'walter.orsi' walter.orsi@teknors.com

Ciao Riccardo, allegati i documenti promessi.

Procedura configurazione ADAMFormato protocollo URXX (UR62 e UR29) in configurazione ADAM.

Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Quarantelli

Santini, Paolo Paolo.Santini@sidel.com; Federico Chiesa federicochiesa@maselli.com; Orsi, Walter [EXTERNAL] Walter.Orsi.external@sidel.com; walter.orsi walter.orsi@teknors.com

Ciao Riccardo, ok per il 6. Ci aggiorniamo prima nel caso ci siano variazioni sul tema. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Riccardo Riccardo.Guerra@sidel.com

Matteo Quarantelli matteoquarantelli@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Federico Chiesa federicochiesa@maselli.com; Orsi, Walter [EXTERNAL] Walter.Orsi.external@sidel.com; walter.orsi walter.orsi@teknors.com

Grazie Pietro Confermo quello che ci siamo detti e tu hai scritto Matteo fammi sapere se la sett 19, martedì 06-05, possiamo trovarci in Maselli per il test, il tempo necessario, se fosse necessario tornare anche il giorno dopo non credo sia un problema Il test prevede:

A carico Sidel

Pc con interfaccia HMI della macchina PLC macchina Alimentatore 24Vdc Convertitore USB-R485 (Moxa) [per programmazione ADAM]

Da parte Maselli:

Strumenti acquistati con il progetto Cavi ….

Test:

Collegamento strumenti e verifica funzionale con il PLC Programmazione/settaggio strumenti, manuale Aggiornamento firmware Adam, manuale

Ciao Riccardo

General

Bracchi pietrobracchi@maselli.com

Federico Chiesa federicochiesa@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Matteo Quarantelli matteoquarantelli@maselli.com

Ciao Riccardo, come appena discusso al telefono, procediamo quindi per la messa a disposizione di tutto il materiale per il 6/05. @Federico: appena rientra, Alessandra modificherà le date in CDO per allinearle tutte al 6/05. Il materiale non andrà spedito, ma reso disponibile per il test che verrà effettuato presso di noi (finito il test, la Sidel poi lo ritirerà) @Riccardo: per la visita da noi per testare insieme il sistema, puoi coordinarti direttamente con Matteo, sulla base di questa data. Grazie a tutti, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Bracchi

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Matteo Quarantelli matteoquarantelli@maselli.com; Federico Chiesa federicochiesa@maselli.com

Ciao Riccardo, scusa per il ritardo nella mia risposta, ma stavamo aspettando anche delle informazioni da un fornitore. Ti confermo che:

La consegna del materiale sotto-riportato è prevista per il 28/04 UR62 A0840184-CA13AA14ADF6N A0840192-BC11BA11UR29 A0840268-AA14LD21D con modulo ADAM A0117271Nota: dal momento che il deflettore DF15 A0840316-ABD1AA91B è già stato acquistato dal cliente finale, non credo abbiate intenzione di prenderlo, ma per favore confermami

La consegna del materiale sotto-riportato è prevista per il 06/05 Kit Conducibilità A0559015Nota: la settimana prossima potremmo avere un eventuale aggiornamento su un possibile anticipo.

Se per te va bene, ci possiamo sentire alla fine della settimana prossima, per capire quale potrebbe essere lo schedule reale per lattività di prova congiunta presso la nostra sede che hai ipotizzato con Matteo. Fammi sapere. Grazie in anticipo. Approfitto di questa mail per chiederti un aggiornamento sul vostro ordine: sai quando dovrebbe arrivare? Ancora grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Pietro Mi sai dire, in base ai vostri piani di produzione, quando riuscite ad avere i componenti per poterli provare insieme a voi. Grazie Ciao Riccardo

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ok grazie Mi torna adesso anche con la richiesta di Ugo Bini Ciao Riccardo

Bracchi pietrobracchi@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Riccardo, anche per lUR62 viene fornito il modulo ADAM; non appare in offerta come riga separata perché linformazione è intrinseca nel modello dellUR62 stesso. Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Grazie Pietro Ho un dubbio soltanto per lUR62 Questo strumento esce direttamente in 4-20mA, giusto? Non ha bisogno del convertitore ADAM come lUR29, lo chiedo perché nello schema condiviso da Matteo ha messo 2 ADAM uno ciascun refrattometro.

Magari mi sono perso qualcosa Grazie Ciao Riccardo

General

Bracchi pietrobracchi@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Riccardo, ti confermo che abbiamo preso in carico la tua richiesta di mettere in produzione, prima dellarrivo del vostro ordine ufficiale, il materiale come da offerte che ri-allego, Appena arriverà il vostro ordine, aggiorneremo la nostra conferma dordine (che per ora non vi condividiamo) con il numero dordine stesso e ve la faremo avere. Nei prossimi giorni ti confermeremo la consegna tentativa, in base alle informazioni che ci verranno dalla Produzione. Allego poi la documentazione disponibile per quanto riguarda il Kit Conducibilità. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Pietro A seguito della telefonata ti scrivo per chiederti se potete mettere in produzione gli strumenti previsti in questo progetto, Diet As Regular - San Giorgio in Bosco -.

Refrattometro UR62 con convertitore ADAM 4-20mA Refrattometro UR29 con convertitore ADAM 4-20mA Conduttivimetro per misura su acqua Ti chiedo anche un manuale del conduttivimetro per poterlo codificare correttamente ed avere unidea della configurazione che poi quando verremo a testare lo spulceremo bene. Metto in copia anche Paolo per la codifica del convertitore ADAM

Attendo tua conferma il manualetto e il tempo di consegna degli strumenti così pianifichiamo nel dettaglio Grazie Ciao Riccardo General General

Bracchi pietrobracchi@maselli.com

Buongiorno Paolo, in allegato la documentazione richiesta relativa alle due offerte nuovamente allegate, per quanto riguarda lUR62 e lUR29 Essendo materiale dacquisto, non abbiamo invece i disegni relativi al conduttivimetro. Resto a disposizione per ulteriori chiarimenti. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Bracchi

Cantarelli, Paolo Paolo.Cantarelli@sidel.com

Buongiorno Paolo, ho chiesto allufficio Tecnico di verificare quali disegni abbiamo disponibili. Appena mi confermano, la informo. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Bolzoni alessandrabolzoni@maselli.com

Buongiorno Paolo, Le risponderà il collega Pietro Bracchi che sta seguendo il progetto e che ci sta leggendo in copia.

Grazie Cordiali saluti Alessandra

Paolo Paolo.Cantarelli@sidel.com

Buongiorno Alessandra Nel ringraziarla per lofferta inviata, le chiedo cortesemente se potrebbe inviarmi anche la documentazione tecnica ed i disegni / modelli 3D dei componenti sottoelencati affinché possa procedere con la codifica

UR62 e relativi attacchi UR29 e relativi attacchi conduttivimetro e relativi attacchi Grazie Cordiali saluti Paolo Cantarelli Mechanical Engineer - Service Engineering

Sidel s.p.a. Via la Spezia 241/A 43126 Parma, ITALY Tel: +39 0521 999566 www.sidel.com P Please consider the environment before printing this e-mail

General

Bolzoni alessandrabolzoni@maselli.com

Cantarelli, Paolo Paolo.Cantarelli@sidel.com; walter.orsi walter.orsi@teknors.com; Pietro Bracchi pietrobracchi@maselli.com; Matteo Quarantelli matteoquarantelli@maselli.com

di test latenza, offerta RDO250201 per UR29 con modulo Adam e conducibilità

Buongiorno, Come da accordi intercorsi, invio la nostra offerta RDO250201 per:

UR29 solo modulo ADAM (e attacco DN40 a saldare)Kit conducibilità composto da sonda e ricevitore (e nocciolo per installazione). Tempo di consegna: 6 settimane dallordine con possibilità di anticipo in caso di esigenze particolari.

Cordiali saluti

Alessandra Bolzoni

Sales coordinator

Maselli Misure S.p.A. P:

  • 39 0521 257400 E: Parma, Italy

www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Bracchi pietrobracchi@maselli.com

Matteo Quarantelli matteoquarantelli@maselli.com

Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; 'walter.orsi' walter.orsi@teknors.com; 'Passera, Alessandro' Alessandro.Passera@sidel.com; 'Cantarelli, Paolo' Paolo.Cantarelli@sidel.com; Alessandra Bolzoni alessandrabolzoni@maselli.com

Questa volta con lallegato. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Bracchi

Matteo Quarantelli matteoquarantelli@maselli.com

Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com; Cantarelli, Paolo Paolo.Cantarelli@sidel.com; Alessandra Bolzoni alessandrabolzoni@maselli.com

Ciao Riccardo, per quanto riguarda la prima riga (UR62, vedi parte evidenziata in giallo sotto), puoi fare riferimento allofferta allegata. Come a tua telefonata con Matteo, a breve ti faremo poi avere lofferta per:

UR29 solo modulo ADAM (e attacco DN40 a saldare)Kit conducibilità composto da sonda e ricevitore (e nocciolo per installazione). Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com; Cantarelli, Paolo Paolo.Cantarelli@sidel.com

Grazie Matteo Ho provato a chiamarti Intendevo dire di tenere i due Adam sui refrattometri, la loro uscita 4-20mA, andrebbe allMP06, lMP06 sarebbe collegato in Profibus ed anche in 4-20mA, verso il PLC, questo per garantire la velocità di trasmissione dai refrattometri, se lMP06 non inserisce latenza nel loop 4-20mA, ed il collegamento Profibus, con lopportuno telegramma vi consentirebbe di ricevere i dati richiesti, ed avremmo un test anche della latenza Profibus su questa macchina. Pietro quando puoi inviarci lofferta aggiornata? Inse4risco nel loop anche Paolo Cantarelli che si occuperà della parte meccanica del progetto, inclusa la codifica dei componenti Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, vedi sotto le risposte. A presto Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Grazie Matteo Bene, lattuatore di che tipo è? Nei 900ms ce anche il suo ritardo di attuazione, anche il tempo di ciclo del PLC ma con una 1200 immagino sia del ms. la sola latenza del loop dellUR potrebbe essere confrontabile con i 300ms attesi. Questo poi lo verifichiamo in campo. OK Direi che sulla prima applicazione verificheremo le prestazioni in campo, ma non mi aspetto problemi anche perché la taglia della macchina non è grande e quindi i flussi sono relativamente bassi. Matteo mi puoi aggiornare lo schema completo? Vedi allegato SLIDE 2 Ha senso inserire comunque lMP06 nel loop di corrente tra i sensori ed il PLC, ingresso 4-20 ed uscita 4-20, si avrebbe un incremento di latenza, non credo sia eccessivo, ma avremmo il vantaggio di poter inviare i dati che avevamo condiviso inizialmente: Intendi al posto dellADAM ? direi di no altrimenti introdurremmo un riardo di alcuni secondi. Se vuoi possiamo predisporre anche lMP06 (SLIDE 1) per una seconda tornata di test. Il test della vita deve essere fatto con gli ADAM (SLIDE 2) ed il log dovrebbe farlo il plc.

centralina MP06 Maselli per gestione Maselli da remoto, alla quale vanno passati alcuni dati: da Sidel a Maselli Alcuni di questi non sono al momento trasferibili da PLC a MP06 (manca telegramma e implementazione sw)

lettura brix UR62 in ingresso linea sciroppo [°Brix] lettura brix UR29 da stream sciroppo [°Brix] temperatura sciroppo [°C] flusso sciroppo [kg/min] flusso acqua [l/min] densità sciroppo [kg/dm3] livello tank sciroppo [%] volume attivo tank sciroppo [L] lettura brix prodotto [°Brix] temperatura prodotto [°C] CO2 prodotto lettura conducibilità dellacqua in ingresso macchina (?) temperatura acqua in ingresso macchina (?) Pressione ingresso linea sciroppo Pressione stream sciroppo

segnali da Maselli a Sidel (Profibus):

Brix sciroppo ingresso linea sciroppo UR62Brix sciroppo stream sciroppo UR29Conducibilità acqua ingressoBrix acqua

Per il resto ti risponde Pietro Chiedo a Pietro se mi può preparare lofferta completa:

UR62 con convertitore RS485/(4-20mA) + attacchi UR29 con convertitore RS485/(4-20mA) + attacchi conduttivimetro con centralina + attacchi vi chiedo inoltre anche i manuali dei sensori, i disegni con le dimensioni, e possibilmente i 3D, tutto questo come sapete per codificare i componenti. Teneteci aggiornati sugli sviluppi del nuovo bus, più veloce, per la comunicazione con gli strumenti. Matteo hai della documentazione oppure è ancora presto? Presto, presto Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, siamo finalmente riusciti a terminare i test con UR29 Di seguito i risultati. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Quarantelli

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Stiamo completando i test .. purtroppo, è stato necessario più tempo del previsto. Spero di avere news domani. Ti tengo aggiornato Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Matteo Dammi buone notizie… Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ok, grazie. Entro domani dovremmo avere il feedback sui test . Ti faccio sapere Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Matteo Ripeto quello che ho detto andava evidenziato nel momento in cui è emersa la questione, lanno scorso, quando è stato deciso di inserire lopzione profibus ed uscite 4-20mA con la MP06. Certo che non voglio fare verifiche di timing nel progetto di San Giorgio. Ho riassunto uno specchietto per definire il perimetro di cosa stiamo parlando.

Machine size 6 9 10
Prod flow Nom [l/h] 27000 54000 60000
Prod flow Min [l/h] 13500 27000 30000
Prod flow Nom [l/s] 7,50 15,00 16,67
Prod flow min [l/s] 3,75 7,50 8,33
300ms Lt eq @Nom speed 2,25 4,5 5
300ms Lt eq @Min speed 1,125 2,25 2,5

Appena ci vediamo ne parliamo meglio. Dai dati che ho lo strumento in Tolosa UR24, fornisce la stessa misura/valore per almeno 2 secondi, come vedi su macchine molto grandi lindeterminazione sui litri di sciroppo diventano troppi. Fammi sapere per i test Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, purtroppo, con MP06 connesso a 2 sensori è praticamente impossibile soddisfare il requisito di 300 ms nemmeno aumentando la velocità della seriale a 115200 (adesso andiamo a 19200) E un problema intrinseco del protocollo master-slave tra mp06 e sensori. Resta in realtà da verificare se questo requisito, allinterno dellintera catena di pilotaggio sia davvero necessario. Noi potremmo, come anticipato, predisporre anche lMP06 per i test comparativi. Cercheremo di fare i test mercoledì o giovedì Ti faccio sapere Grazie Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Grazie Matteo Fammi sapere quando riuscite ad effettuare il test “latenza sensore con convertitore RS485 / 4..20 mA”. Sono sicuro che la tempistica dei 300ms, è emersa sin dai primi incontri in Maselli, è per questa ragione che era stato deciso di fornire lMP06 con le uscite in corrente e Profibus, per poi verificare in cantiere se il Profibus fosse sufficiente pensando alle future applicazioni. Una domanda che mi viene spontanea è possibile aumentare il più possibile il baudrate della seriale? Dato che è una 485 differenziale e la lunghezza del cavo è minima. Gradirei partecipare al test, se possibile. In quelloccasione potremmo vedere nel dettaglio i dati raccolti nel mixer di Tolosa, il primo mixer della nuova generazione sidel, fornito di un UR24 con convertitore Profibus Installato nel seguente modo:

Ricordo, inoltre per la cronaca, che al mio rientro da Tolosa diciamo nel 2012, sono passato in Maselli e parlando di questo argomento con Paolo (il softwarista in “pensione”) mi diceva che questa “scarsa” velocità dellUR24 sarebbe stata risolta, oppure gestibile con la nuova generazione di strumenti equipaggiati con un nuovo microcontrollore. Fatemi sapere, Andrea M. fammi sapere per laltro tema a riguardo degli strumenti da inserire al “volo” per avere dati sui trend del brix, prima del progetto. Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com

Ciao riccardo ecco le 2 slides. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

Adjuntos


  • a1b9e7335b6848cfee9d638cd2e5229d

{EXT} utility

  • 06-05-2025

MAU_V_2_9_6.msi

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|


  • 464fdf3553e3cc9fdf3681f912bcf1e2

{EXT} I: foto procedure adam sidel

  • 06-05-2025

foto procedure modulo adam Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Adjuntos

  • 04-SIDEL/09 - SAE452 - Diet as Regular - San Giovanni in Bosco/adjuntos/Screenshot 2025-05-06 111701.jpg
  • 04-SIDEL/09 - SAE452 - Diet as Regular - San Giovanni in Bosco/adjuntos/Screenshot 2025-05-06 111755.jpg
  • 04-SIDEL/09 - SAE452 - Diet as Regular - San Giovanni in Bosco/adjuntos/Screenshot 2025-05-06 113243.jpg
  • 04-SIDEL/09 - SAE452 - Diet as Regular - San Giovanni in Bosco/adjuntos/Screenshot 2025-05-06 114159.jpg

  • 4fce380d128f4277f302ff9ea1b8946b

{EXT} R: SIDEL (01011476) R: Nuova configurazione e banco di test latenza, offerta RDO250201 per UR29 con modulo Adam e conducibilità

  • 06-05-2025

Invio versione fw ADAM con nota nella procedura v3 Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Quarantelli

'Orsi, Walter [EXTERNAL]' Walter.Orsi.external@sidel.com; 'walter.orsi' walter.orsi@teknors.com

Inoltro documento aggiornato (V2) Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Quarantelli

'Santini, Paolo' Paolo.Santini@sidel.com; Federico Chiesa federicochiesa@maselli.com; 'Orsi, Walter [EXTERNAL]' Walter.Orsi.external@sidel.com; 'walter.orsi' walter.orsi@teknors.com

Ciao Riccardo, allegati i documenti promessi.

Procedura configurazione ADAMFormato protocollo URXX (UR62 e UR29) in configurazione ADAM.

Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Quarantelli

Santini, Paolo Paolo.Santini@sidel.com; Federico Chiesa federicochiesa@maselli.com; Orsi, Walter [EXTERNAL] Walter.Orsi.external@sidel.com; walter.orsi walter.orsi@teknors.com

Ciao Riccardo, ok per il 6. Ci aggiorniamo prima nel caso ci siano variazioni sul tema. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Riccardo Riccardo.Guerra@sidel.com

Matteo Quarantelli matteoquarantelli@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Federico Chiesa federicochiesa@maselli.com; Orsi, Walter [EXTERNAL] Walter.Orsi.external@sidel.com; walter.orsi walter.orsi@teknors.com

Grazie Pietro Confermo quello che ci siamo detti e tu hai scritto Matteo fammi sapere se la sett 19, martedì 06-05, possiamo trovarci in Maselli per il test, il tempo necessario, se fosse necessario tornare anche il giorno dopo non credo sia un problema Il test prevede:

A carico Sidel

Pc con interfaccia HMI della macchina PLC macchina Alimentatore 24Vdc Convertitore USB-R485 (Moxa) [per programmazione ADAM]

Da parte Maselli:

Strumenti acquistati con il progetto Cavi ….

Test:

Collegamento strumenti e verifica funzionale con il PLC Programmazione/settaggio strumenti, manuale Aggiornamento firmware Adam, manuale

Ciao Riccardo

General

Bracchi pietrobracchi@maselli.com

Federico Chiesa federicochiesa@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Matteo Quarantelli matteoquarantelli@maselli.com

Ciao Riccardo, come appena discusso al telefono, procediamo quindi per la messa a disposizione di tutto il materiale per il 6/05. @Federico: appena rientra, Alessandra modificherà le date in CDO per allinearle tutte al 6/05. Il materiale non andrà spedito, ma reso disponibile per il test che verrà effettuato presso di noi (finito il test, la Sidel poi lo ritirerà) @Riccardo: per la visita da noi per testare insieme il sistema, puoi coordinarti direttamente con Matteo, sulla base di questa data. Grazie a tutti, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Bracchi

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Matteo Quarantelli matteoquarantelli@maselli.com; Federico Chiesa federicochiesa@maselli.com

Ciao Riccardo, scusa per il ritardo nella mia risposta, ma stavamo aspettando anche delle informazioni da un fornitore. Ti confermo che:

La consegna del materiale sotto-riportato è prevista per il 28/04 UR62 A0840184-CA13AA14ADF6N A0840192-BC11BA11UR29 A0840268-AA14LD21D con modulo ADAM A0117271Nota: dal momento che il deflettore DF15 A0840316-ABD1AA91B è già stato acquistato dal cliente finale, non credo abbiate intenzione di prenderlo, ma per favore confermami

La consegna del materiale sotto-riportato è prevista per il 06/05 Kit Conducibilità A0559015Nota: la settimana prossima potremmo avere un eventuale aggiornamento su un possibile anticipo.

Se per te va bene, ci possiamo sentire alla fine della settimana prossima, per capire quale potrebbe essere lo schedule reale per lattività di prova congiunta presso la nostra sede che hai ipotizzato con Matteo. Fammi sapere. Grazie in anticipo. Approfitto di questa mail per chiederti un aggiornamento sul vostro ordine: sai quando dovrebbe arrivare? Ancora grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Pietro Mi sai dire, in base ai vostri piani di produzione, quando riuscite ad avere i componenti per poterli provare insieme a voi. Grazie Ciao Riccardo

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ok grazie Mi torna adesso anche con la richiesta di Ugo Bini Ciao Riccardo

Bracchi pietrobracchi@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Riccardo, anche per lUR62 viene fornito il modulo ADAM; non appare in offerta come riga separata perché linformazione è intrinseca nel modello dellUR62 stesso. Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Grazie Pietro Ho un dubbio soltanto per lUR62 Questo strumento esce direttamente in 4-20mA, giusto? Non ha bisogno del convertitore ADAM come lUR29, lo chiedo perché nello schema condiviso da Matteo ha messo 2 ADAM uno ciascun refrattometro.

Magari mi sono perso qualcosa Grazie Ciao Riccardo

General

Bracchi pietrobracchi@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Riccardo, ti confermo che abbiamo preso in carico la tua richiesta di mettere in produzione, prima dellarrivo del vostro ordine ufficiale, il materiale come da offerte che ri-allego, Appena arriverà il vostro ordine, aggiorneremo la nostra conferma dordine (che per ora non vi condividiamo) con il numero dordine stesso e ve la faremo avere. Nei prossimi giorni ti confermeremo la consegna tentativa, in base alle informazioni che ci verranno dalla Produzione. Allego poi la documentazione disponibile per quanto riguarda il Kit Conducibilità. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Pietro A seguito della telefonata ti scrivo per chiederti se potete mettere in produzione gli strumenti previsti in questo progetto, Diet As Regular - San Giorgio in Bosco -.

Refrattometro UR62 con convertitore ADAM 4-20mA Refrattometro UR29 con convertitore ADAM 4-20mA Conduttivimetro per misura su acqua Ti chiedo anche un manuale del conduttivimetro per poterlo codificare correttamente ed avere unidea della configurazione che poi quando verremo a testare lo spulceremo bene. Metto in copia anche Paolo per la codifica del convertitore ADAM

Attendo tua conferma il manualetto e il tempo di consegna degli strumenti così pianifichiamo nel dettaglio Grazie Ciao Riccardo General General

Bracchi pietrobracchi@maselli.com

Buongiorno Paolo, in allegato la documentazione richiesta relativa alle due offerte nuovamente allegate, per quanto riguarda lUR62 e lUR29 Essendo materiale dacquisto, non abbiamo invece i disegni relativi al conduttivimetro. Resto a disposizione per ulteriori chiarimenti. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Bracchi

Cantarelli, Paolo Paolo.Cantarelli@sidel.com

Buongiorno Paolo, ho chiesto allufficio Tecnico di verificare quali disegni abbiamo disponibili. Appena mi confermano, la informo. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Bolzoni alessandrabolzoni@maselli.com

Buongiorno Paolo, Le risponderà il collega Pietro Bracchi che sta seguendo il progetto e che ci sta leggendo in copia.

Grazie Cordiali saluti Alessandra

Paolo Paolo.Cantarelli@sidel.com

Buongiorno Alessandra Nel ringraziarla per lofferta inviata, le chiedo cortesemente se potrebbe inviarmi anche la documentazione tecnica ed i disegni / modelli 3D dei componenti sottoelencati affinché possa procedere con la codifica

UR62 e relativi attacchi UR29 e relativi attacchi conduttivimetro e relativi attacchi Grazie Cordiali saluti Paolo Cantarelli Mechanical Engineer - Service Engineering

Sidel s.p.a. Via la Spezia 241/A 43126 Parma, ITALY Tel: +39 0521 999566 www.sidel.com P Please consider the environment before printing this e-mail

General

Bolzoni alessandrabolzoni@maselli.com

Cantarelli, Paolo Paolo.Cantarelli@sidel.com; walter.orsi walter.orsi@teknors.com; Pietro Bracchi pietrobracchi@maselli.com; Matteo Quarantelli matteoquarantelli@maselli.com

di test latenza, offerta RDO250201 per UR29 con modulo Adam e conducibilità

Buongiorno, Come da accordi intercorsi, invio la nostra offerta RDO250201 per:

UR29 solo modulo ADAM (e attacco DN40 a saldare)Kit conducibilità composto da sonda e ricevitore (e nocciolo per installazione). Tempo di consegna: 6 settimane dallordine con possibilità di anticipo in caso di esigenze particolari.

Cordiali saluti

Alessandra Bolzoni

Sales coordinator

Maselli Misure S.p.A. P:

  • 39 0521 257400 E: Parma, Italy

www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Bracchi pietrobracchi@maselli.com

Matteo Quarantelli matteoquarantelli@maselli.com

Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; 'walter.orsi' walter.orsi@teknors.com; 'Passera, Alessandro' Alessandro.Passera@sidel.com; 'Cantarelli, Paolo' Paolo.Cantarelli@sidel.com; Alessandra Bolzoni alessandrabolzoni@maselli.com

Questa volta con lallegato. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Bracchi

Matteo Quarantelli matteoquarantelli@maselli.com

Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com; Cantarelli, Paolo Paolo.Cantarelli@sidel.com; Alessandra Bolzoni alessandrabolzoni@maselli.com

Ciao Riccardo, per quanto riguarda la prima riga (UR62, vedi parte evidenziata in giallo sotto), puoi fare riferimento allofferta allegata. Come a tua telefonata con Matteo, a breve ti faremo poi avere lofferta per:

UR29 solo modulo ADAM (e attacco DN40 a saldare)Kit conducibilità composto da sonda e ricevitore (e nocciolo per installazione). Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com; Cantarelli, Paolo Paolo.Cantarelli@sidel.com

Grazie Matteo Ho provato a chiamarti Intendevo dire di tenere i due Adam sui refrattometri, la loro uscita 4-20mA, andrebbe allMP06, lMP06 sarebbe collegato in Profibus ed anche in 4-20mA, verso il PLC, questo per garantire la velocità di trasmissione dai refrattometri, se lMP06 non inserisce latenza nel loop 4-20mA, ed il collegamento Profibus, con lopportuno telegramma vi consentirebbe di ricevere i dati richiesti, ed avremmo un test anche della latenza Profibus su questa macchina. Pietro quando puoi inviarci lofferta aggiornata? Inse4risco nel loop anche Paolo Cantarelli che si occuperà della parte meccanica del progetto, inclusa la codifica dei componenti Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, vedi sotto le risposte. A presto Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Grazie Matteo Bene, lattuatore di che tipo è? Nei 900ms ce anche il suo ritardo di attuazione, anche il tempo di ciclo del PLC ma con una 1200 immagino sia del ms. la sola latenza del loop dellUR potrebbe essere confrontabile con i 300ms attesi. Questo poi lo verifichiamo in campo. OK Direi che sulla prima applicazione verificheremo le prestazioni in campo, ma non mi aspetto problemi anche perché la taglia della macchina non è grande e quindi i flussi sono relativamente bassi. Matteo mi puoi aggiornare lo schema completo? Vedi allegato SLIDE 2 Ha senso inserire comunque lMP06 nel loop di corrente tra i sensori ed il PLC, ingresso 4-20 ed uscita 4-20, si avrebbe un incremento di latenza, non credo sia eccessivo, ma avremmo il vantaggio di poter inviare i dati che avevamo condiviso inizialmente: Intendi al posto dellADAM ? direi di no altrimenti introdurremmo un riardo di alcuni secondi. Se vuoi possiamo predisporre anche lMP06 (SLIDE 1) per una seconda tornata di test. Il test della vita deve essere fatto con gli ADAM (SLIDE 2) ed il log dovrebbe farlo il plc.

centralina MP06 Maselli per gestione Maselli da remoto, alla quale vanno passati alcuni dati: da Sidel a Maselli Alcuni di questi non sono al momento trasferibili da PLC a MP06 (manca telegramma e implementazione sw)

lettura brix UR62 in ingresso linea sciroppo [°Brix] lettura brix UR29 da stream sciroppo [°Brix] temperatura sciroppo [°C] flusso sciroppo [kg/min] flusso acqua [l/min] densità sciroppo [kg/dm3] livello tank sciroppo [%] volume attivo tank sciroppo [L] lettura brix prodotto [°Brix] temperatura prodotto [°C] CO2 prodotto lettura conducibilità dellacqua in ingresso macchina (?) temperatura acqua in ingresso macchina (?) Pressione ingresso linea sciroppo Pressione stream sciroppo

segnali da Maselli a Sidel (Profibus):

Brix sciroppo ingresso linea sciroppo UR62Brix sciroppo stream sciroppo UR29Conducibilità acqua ingressoBrix acqua

Per il resto ti risponde Pietro Chiedo a Pietro se mi può preparare lofferta completa:

UR62 con convertitore RS485/(4-20mA) + attacchi UR29 con convertitore RS485/(4-20mA) + attacchi conduttivimetro con centralina + attacchi vi chiedo inoltre anche i manuali dei sensori, i disegni con le dimensioni, e possibilmente i 3D, tutto questo come sapete per codificare i componenti. Teneteci aggiornati sugli sviluppi del nuovo bus, più veloce, per la comunicazione con gli strumenti. Matteo hai della documentazione oppure è ancora presto? Presto, presto Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, siamo finalmente riusciti a terminare i test con UR29 Di seguito i risultati. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Quarantelli

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Stiamo completando i test .. purtroppo, è stato necessario più tempo del previsto. Spero di avere news domani. Ti tengo aggiornato Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Matteo Dammi buone notizie… Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ok, grazie. Entro domani dovremmo avere il feedback sui test . Ti faccio sapere Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Matteo Ripeto quello che ho detto andava evidenziato nel momento in cui è emersa la questione, lanno scorso, quando è stato deciso di inserire lopzione profibus ed uscite 4-20mA con la MP06. Certo che non voglio fare verifiche di timing nel progetto di San Giorgio. Ho riassunto uno specchietto per definire il perimetro di cosa stiamo parlando.

Machine size 6 9 10
Prod flow Nom [l/h] 27000 54000 60000
Prod flow Min [l/h] 13500 27000 30000
Prod flow Nom [l/s] 7,50 15,00 16,67
Prod flow min [l/s] 3,75 7,50 8,33
300ms Lt eq @Nom speed 2,25 4,5 5
300ms Lt eq @Min speed 1,125 2,25 2,5

Appena ci vediamo ne parliamo meglio. Dai dati che ho lo strumento in Tolosa UR24, fornisce la stessa misura/valore per almeno 2 secondi, come vedi su macchine molto grandi lindeterminazione sui litri di sciroppo diventano troppi. Fammi sapere per i test Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, purtroppo, con MP06 connesso a 2 sensori è praticamente impossibile soddisfare il requisito di 300 ms nemmeno aumentando la velocità della seriale a 115200 (adesso andiamo a 19200) E un problema intrinseco del protocollo master-slave tra mp06 e sensori. Resta in realtà da verificare se questo requisito, allinterno dellintera catena di pilotaggio sia davvero necessario. Noi potremmo, come anticipato, predisporre anche lMP06 per i test comparativi. Cercheremo di fare i test mercoledì o giovedì Ti faccio sapere Grazie Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Grazie Matteo Fammi sapere quando riuscite ad effettuare il test “latenza sensore con convertitore RS485 / 4..20 mA”. Sono sicuro che la tempistica dei 300ms, è emersa sin dai primi incontri in Maselli, è per questa ragione che era stato deciso di fornire lMP06 con le uscite in corrente e Profibus, per poi verificare in cantiere se il Profibus fosse sufficiente pensando alle future applicazioni. Una domanda che mi viene spontanea è possibile aumentare il più possibile il baudrate della seriale? Dato che è una 485 differenziale e la lunghezza del cavo è minima. Gradirei partecipare al test, se possibile. In quelloccasione potremmo vedere nel dettaglio i dati raccolti nel mixer di Tolosa, il primo mixer della nuova generazione sidel, fornito di un UR24 con convertitore Profibus Installato nel seguente modo:

Ricordo, inoltre per la cronaca, che al mio rientro da Tolosa diciamo nel 2012, sono passato in Maselli e parlando di questo argomento con Paolo (il softwarista in “pensione”) mi diceva che questa “scarsa” velocità dellUR24 sarebbe stata risolta, oppure gestibile con la nuova generazione di strumenti equipaggiati con un nuovo microcontrollore. Fatemi sapere, Andrea M. fammi sapere per laltro tema a riguardo degli strumenti da inserire al “volo” per avere dati sui trend del brix, prima del progetto. Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com

Ciao riccardo ecco le 2 slides. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

Adjuntos


  • 34a87ed5874707d5bb51a2785d6dfbd6

{EXT} R: SIDEL (01011476) R: Nuova configurazione e banco di test latenza, offerta RDO250201 per UR29 con modulo Adam e conducibilità

  • 06-05-2025

Inoltro documento aggiornato (V2) Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Quarantelli

'Orsi, Walter [EXTERNAL]' Walter.Orsi.external@sidel.com; 'walter.orsi' walter.orsi@teknors.com

Ciao Riccardo, allegati i documenti promessi.

Procedura configurazione ADAMFormato protocollo URXX (UR62 e UR29) in configurazione ADAM.

Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Quarantelli

Santini, Paolo Paolo.Santini@sidel.com; Federico Chiesa federicochiesa@maselli.com; Orsi, Walter [EXTERNAL] Walter.Orsi.external@sidel.com; walter.orsi walter.orsi@teknors.com

Ciao Riccardo, ok per il 6. Ci aggiorniamo prima nel caso ci siano variazioni sul tema. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Riccardo Riccardo.Guerra@sidel.com

Matteo Quarantelli matteoquarantelli@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Federico Chiesa federicochiesa@maselli.com; Orsi, Walter [EXTERNAL] Walter.Orsi.external@sidel.com; walter.orsi walter.orsi@teknors.com

Grazie Pietro Confermo quello che ci siamo detti e tu hai scritto Matteo fammi sapere se la sett 19, martedì 06-05, possiamo trovarci in Maselli per il test, il tempo necessario, se fosse necessario tornare anche il giorno dopo non credo sia un problema Il test prevede:

A carico Sidel

Pc con interfaccia HMI della macchina PLC macchina Alimentatore 24Vdc Convertitore USB-R485 (Moxa) [per programmazione ADAM]

Da parte Maselli:

Strumenti acquistati con il progetto Cavi ….

Test:

Collegamento strumenti e verifica funzionale con il PLC Programmazione/settaggio strumenti, manuale Aggiornamento firmware Adam, manuale

Ciao Riccardo

General

Bracchi pietrobracchi@maselli.com

Federico Chiesa federicochiesa@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Matteo Quarantelli matteoquarantelli@maselli.com

Ciao Riccardo, come appena discusso al telefono, procediamo quindi per la messa a disposizione di tutto il materiale per il 6/05. @Federico: appena rientra, Alessandra modificherà le date in CDO per allinearle tutte al 6/05. Il materiale non andrà spedito, ma reso disponibile per il test che verrà effettuato presso di noi (finito il test, la Sidel poi lo ritirerà) @Riccardo: per la visita da noi per testare insieme il sistema, puoi coordinarti direttamente con Matteo, sulla base di questa data. Grazie a tutti, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Bracchi

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com; Matteo Quarantelli matteoquarantelli@maselli.com; Federico Chiesa federicochiesa@maselli.com

Ciao Riccardo, scusa per il ritardo nella mia risposta, ma stavamo aspettando anche delle informazioni da un fornitore. Ti confermo che:

La consegna del materiale sotto-riportato è prevista per il 28/04 UR62 A0840184-CA13AA14ADF6N A0840192-BC11BA11UR29 A0840268-AA14LD21D con modulo ADAM A0117271Nota: dal momento che il deflettore DF15 A0840316-ABD1AA91B è già stato acquistato dal cliente finale, non credo abbiate intenzione di prenderlo, ma per favore confermami

La consegna del materiale sotto-riportato è prevista per il 06/05 Kit Conducibilità A0559015Nota: la settimana prossima potremmo avere un eventuale aggiornamento su un possibile anticipo.

Se per te va bene, ci possiamo sentire alla fine della settimana prossima, per capire quale potrebbe essere lo schedule reale per lattività di prova congiunta presso la nostra sede che hai ipotizzato con Matteo. Fammi sapere. Grazie in anticipo. Approfitto di questa mail per chiederti un aggiornamento sul vostro ordine: sai quando dovrebbe arrivare? Ancora grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Pietro Mi sai dire, in base ai vostri piani di produzione, quando riuscite ad avere i componenti per poterli provare insieme a voi. Grazie Ciao Riccardo

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ok grazie Mi torna adesso anche con la richiesta di Ugo Bini Ciao Riccardo

Bracchi pietrobracchi@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Riccardo, anche per lUR62 viene fornito il modulo ADAM; non appare in offerta come riga separata perché linformazione è intrinseca nel modello dellUR62 stesso. Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Grazie Pietro Ho un dubbio soltanto per lUR62 Questo strumento esce direttamente in 4-20mA, giusto? Non ha bisogno del convertitore ADAM come lUR29, lo chiedo perché nello schema condiviso da Matteo ha messo 2 ADAM uno ciascun refrattometro.

Magari mi sono perso qualcosa Grazie Ciao Riccardo

General

Bracchi pietrobracchi@maselli.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Riccardo, ti confermo che abbiamo preso in carico la tua richiesta di mettere in produzione, prima dellarrivo del vostro ordine ufficiale, il materiale come da offerte che ri-allego, Appena arriverà il vostro ordine, aggiorneremo la nostra conferma dordine (che per ora non vi condividiamo) con il numero dordine stesso e ve la faremo avere. Nei prossimi giorni ti confermeremo la consegna tentativa, in base alle informazioni che ci verranno dalla Produzione. Allego poi la documentazione disponibile per quanto riguarda il Kit Conducibilità. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Pietro A seguito della telefonata ti scrivo per chiederti se potete mettere in produzione gli strumenti previsti in questo progetto, Diet As Regular - San Giorgio in Bosco -.

Refrattometro UR62 con convertitore ADAM 4-20mA Refrattometro UR29 con convertitore ADAM 4-20mA Conduttivimetro per misura su acqua Ti chiedo anche un manuale del conduttivimetro per poterlo codificare correttamente ed avere unidea della configurazione che poi quando verremo a testare lo spulceremo bene. Metto in copia anche Paolo per la codifica del convertitore ADAM

Attendo tua conferma il manualetto e il tempo di consegna degli strumenti così pianifichiamo nel dettaglio Grazie Ciao Riccardo General General

Bracchi pietrobracchi@maselli.com

Buongiorno Paolo, in allegato la documentazione richiesta relativa alle due offerte nuovamente allegate, per quanto riguarda lUR62 e lUR29 Essendo materiale dacquisto, non abbiamo invece i disegni relativi al conduttivimetro. Resto a disposizione per ulteriori chiarimenti. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Bracchi

Cantarelli, Paolo Paolo.Cantarelli@sidel.com

Buongiorno Paolo, ho chiesto allufficio Tecnico di verificare quali disegni abbiamo disponibili. Appena mi confermano, la informo. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Bolzoni alessandrabolzoni@maselli.com

Buongiorno Paolo, Le risponderà il collega Pietro Bracchi che sta seguendo il progetto e che ci sta leggendo in copia.

Grazie Cordiali saluti Alessandra

Paolo Paolo.Cantarelli@sidel.com

Buongiorno Alessandra Nel ringraziarla per lofferta inviata, le chiedo cortesemente se potrebbe inviarmi anche la documentazione tecnica ed i disegni / modelli 3D dei componenti sottoelencati affinché possa procedere con la codifica

UR62 e relativi attacchi UR29 e relativi attacchi conduttivimetro e relativi attacchi Grazie Cordiali saluti Paolo Cantarelli Mechanical Engineer - Service Engineering

Sidel s.p.a. Via la Spezia 241/A 43126 Parma, ITALY Tel: +39 0521 999566 www.sidel.com P Please consider the environment before printing this e-mail

General

Bolzoni alessandrabolzoni@maselli.com

Cantarelli, Paolo Paolo.Cantarelli@sidel.com; walter.orsi walter.orsi@teknors.com; Pietro Bracchi pietrobracchi@maselli.com; Matteo Quarantelli matteoquarantelli@maselli.com

di test latenza, offerta RDO250201 per UR29 con modulo Adam e conducibilità

Buongiorno, Come da accordi intercorsi, invio la nostra offerta RDO250201 per:

UR29 solo modulo ADAM (e attacco DN40 a saldare)Kit conducibilità composto da sonda e ricevitore (e nocciolo per installazione). Tempo di consegna: 6 settimane dallordine con possibilità di anticipo in caso di esigenze particolari.

Cordiali saluti

Alessandra Bolzoni

Sales coordinator

Maselli Misure S.p.A. P:

  • 39 0521 257400 E: Parma, Italy

www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Bracchi pietrobracchi@maselli.com

Matteo Quarantelli matteoquarantelli@maselli.com

Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; 'walter.orsi' walter.orsi@teknors.com; 'Passera, Alessandro' Alessandro.Passera@sidel.com; 'Cantarelli, Paolo' Paolo.Cantarelli@sidel.com; Alessandra Bolzoni alessandrabolzoni@maselli.com

Questa volta con lallegato. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Bracchi

Matteo Quarantelli matteoquarantelli@maselli.com

Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com; Cantarelli, Paolo Paolo.Cantarelli@sidel.com; Alessandra Bolzoni alessandrabolzoni@maselli.com

Ciao Riccardo, per quanto riguarda la prima riga (UR62, vedi parte evidenziata in giallo sotto), puoi fare riferimento allofferta allegata. Come a tua telefonata con Matteo, a breve ti faremo poi avere lofferta per:

UR29 solo modulo ADAM (e attacco DN40 a saldare)Kit conducibilità composto da sonda e ricevitore (e nocciolo per installazione). Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com; Cantarelli, Paolo Paolo.Cantarelli@sidel.com

Grazie Matteo Ho provato a chiamarti Intendevo dire di tenere i due Adam sui refrattometri, la loro uscita 4-20mA, andrebbe allMP06, lMP06 sarebbe collegato in Profibus ed anche in 4-20mA, verso il PLC, questo per garantire la velocità di trasmissione dai refrattometri, se lMP06 non inserisce latenza nel loop 4-20mA, ed il collegamento Profibus, con lopportuno telegramma vi consentirebbe di ricevere i dati richiesti, ed avremmo un test anche della latenza Profibus su questa macchina. Pietro quando puoi inviarci lofferta aggiornata? Inse4risco nel loop anche Paolo Cantarelli che si occuperà della parte meccanica del progetto, inclusa la codifica dei componenti Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, vedi sotto le risposte. A presto Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Grazie Matteo Bene, lattuatore di che tipo è? Nei 900ms ce anche il suo ritardo di attuazione, anche il tempo di ciclo del PLC ma con una 1200 immagino sia del ms. la sola latenza del loop dellUR potrebbe essere confrontabile con i 300ms attesi. Questo poi lo verifichiamo in campo. OK Direi che sulla prima applicazione verificheremo le prestazioni in campo, ma non mi aspetto problemi anche perché la taglia della macchina non è grande e quindi i flussi sono relativamente bassi. Matteo mi puoi aggiornare lo schema completo? Vedi allegato SLIDE 2 Ha senso inserire comunque lMP06 nel loop di corrente tra i sensori ed il PLC, ingresso 4-20 ed uscita 4-20, si avrebbe un incremento di latenza, non credo sia eccessivo, ma avremmo il vantaggio di poter inviare i dati che avevamo condiviso inizialmente: Intendi al posto dellADAM ? direi di no altrimenti introdurremmo un riardo di alcuni secondi. Se vuoi possiamo predisporre anche lMP06 (SLIDE 1) per una seconda tornata di test. Il test della vita deve essere fatto con gli ADAM (SLIDE 2) ed il log dovrebbe farlo il plc.

centralina MP06 Maselli per gestione Maselli da remoto, alla quale vanno passati alcuni dati: da Sidel a Maselli Alcuni di questi non sono al momento trasferibili da PLC a MP06 (manca telegramma e implementazione sw)

lettura brix UR62 in ingresso linea sciroppo [°Brix] lettura brix UR29 da stream sciroppo [°Brix] temperatura sciroppo [°C] flusso sciroppo [kg/min] flusso acqua [l/min] densità sciroppo [kg/dm3] livello tank sciroppo [%] volume attivo tank sciroppo [L] lettura brix prodotto [°Brix] temperatura prodotto [°C] CO2 prodotto lettura conducibilità dellacqua in ingresso macchina (?) temperatura acqua in ingresso macchina (?) Pressione ingresso linea sciroppo Pressione stream sciroppo

segnali da Maselli a Sidel (Profibus):

Brix sciroppo ingresso linea sciroppo UR62Brix sciroppo stream sciroppo UR29Conducibilità acqua ingressoBrix acqua

Per il resto ti risponde Pietro Chiedo a Pietro se mi può preparare lofferta completa:

UR62 con convertitore RS485/(4-20mA) + attacchi UR29 con convertitore RS485/(4-20mA) + attacchi conduttivimetro con centralina + attacchi vi chiedo inoltre anche i manuali dei sensori, i disegni con le dimensioni, e possibilmente i 3D, tutto questo come sapete per codificare i componenti. Teneteci aggiornati sugli sviluppi del nuovo bus, più veloce, per la comunicazione con gli strumenti. Matteo hai della documentazione oppure è ancora presto? Presto, presto Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, siamo finalmente riusciti a terminare i test con UR29 Di seguito i risultati. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Quarantelli

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Stiamo completando i test .. purtroppo, è stato necessario più tempo del previsto. Spero di avere news domani. Ti tengo aggiornato Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Matteo Dammi buone notizie… Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ok, grazie. Entro domani dovremmo avere il feedback sui test . Ti faccio sapere Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Matteo Ripeto quello che ho detto andava evidenziato nel momento in cui è emersa la questione, lanno scorso, quando è stato deciso di inserire lopzione profibus ed uscite 4-20mA con la MP06. Certo che non voglio fare verifiche di timing nel progetto di San Giorgio. Ho riassunto uno specchietto per definire il perimetro di cosa stiamo parlando.

Machine size 6 9 10
Prod flow Nom [l/h] 27000 54000 60000
Prod flow Min [l/h] 13500 27000 30000
Prod flow Nom [l/s] 7,50 15,00 16,67
Prod flow min [l/s] 3,75 7,50 8,33
300ms Lt eq @Nom speed 2,25 4,5 5
300ms Lt eq @Min speed 1,125 2,25 2,5

Appena ci vediamo ne parliamo meglio. Dai dati che ho lo strumento in Tolosa UR24, fornisce la stessa misura/valore per almeno 2 secondi, come vedi su macchine molto grandi lindeterminazione sui litri di sciroppo diventano troppi. Fammi sapere per i test Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, purtroppo, con MP06 connesso a 2 sensori è praticamente impossibile soddisfare il requisito di 300 ms nemmeno aumentando la velocità della seriale a 115200 (adesso andiamo a 19200) E un problema intrinseco del protocollo master-slave tra mp06 e sensori. Resta in realtà da verificare se questo requisito, allinterno dellintera catena di pilotaggio sia davvero necessario. Noi potremmo, come anticipato, predisporre anche lMP06 per i test comparativi. Cercheremo di fare i test mercoledì o giovedì Ti faccio sapere Grazie Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Grazie Matteo Fammi sapere quando riuscite ad effettuare il test “latenza sensore con convertitore RS485 / 4..20 mA”. Sono sicuro che la tempistica dei 300ms, è emersa sin dai primi incontri in Maselli, è per questa ragione che era stato deciso di fornire lMP06 con le uscite in corrente e Profibus, per poi verificare in cantiere se il Profibus fosse sufficiente pensando alle future applicazioni. Una domanda che mi viene spontanea è possibile aumentare il più possibile il baudrate della seriale? Dato che è una 485 differenziale e la lunghezza del cavo è minima. Gradirei partecipare al test, se possibile. In quelloccasione potremmo vedere nel dettaglio i dati raccolti nel mixer di Tolosa, il primo mixer della nuova generazione sidel, fornito di un UR24 con convertitore Profibus Installato nel seguente modo:

Ricordo, inoltre per la cronaca, che al mio rientro da Tolosa diciamo nel 2012, sono passato in Maselli e parlando di questo argomento con Paolo (il softwarista in “pensione”) mi diceva che questa “scarsa” velocità dellUR24 sarebbe stata risolta, oppure gestibile con la nuova generazione di strumenti equipaggiati con un nuovo microcontrollore. Fatemi sapere, Andrea M. fammi sapere per laltro tema a riguardo degli strumenti da inserire al “volo” per avere dati sui trend del brix, prima del progetto. Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com

Ciao riccardo ecco le 2 slides. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

This e-mail may contain confidential and/or privileged information. If you are not the intended recipient (or have received this e-mail in error), please notify the sender immediately and destroy this e-mail. Any unauthorized copying, disclosure or distribution of the material in this e-mail is strictly forbidden.

Adjuntos


  • 8c802a79438e84d56eddc01e9f1be210

{EXT} R: SIDEL (01011476) R: Nuova configurazione e banco di test latenza, offerta RDO250201 per UR29 con modulo Adam e conducibilità

  • 15-04-2025

Ciao Riccardo, ok per il 6. Ci aggiorniamo prima nel caso ci siano variazioni sul tema. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Walter [EXTERNAL] Walter.Orsi.external@sidel.com; walter.orsi walter.orsi@teknors.com

Grazie Pietro Confermo quello che ci siamo detti e tu hai scritto Matteo fammi sapere se la sett 19, martedì 06-05, possiamo trovarci in Maselli per il test, il tempo necessario, se fosse necessario tornare anche il giorno dopo non credo sia un problema Il test prevede:

A carico Sidel

Pc con interfaccia HMI della macchina PLC macchina Alimentatore 24Vdc Convertitore USB-R485 (Moxa) [per programmazione ADAM]

Da parte Maselli:

Strumenti acquistati con il progetto Cavi ….

Test:

Collegamento strumenti e verifica funzionale con il PLC Programmazione/settaggio strumenti, manuale Aggiornamento firmware Adam, manuale

Ciao Riccardo

General

Bracchi pietrobracchi@maselli.com

Santini, Paolo Paolo.Santini@sidel.com; Matteo Quarantelli matteoquarantelli@maselli.com

Ciao Riccardo, come appena discusso al telefono, procediamo quindi per la messa a disposizione di tutto il materiale per il 6/05. @Federico: appena rientra, Alessandra modificherà le date in CDO per allinearle tutte al 6/05. Il materiale non andrà spedito, ma reso disponibile per il test che verrà effettuato presso di noi (finito il test, la Sidel poi lo ritirerà) @Riccardo: per la visita da noi per testare insieme il sistema, puoi coordinarti direttamente con Matteo, sulla base di questa data. Grazie a tutti, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Bracchi

Santini, Paolo Paolo.Santini@sidel.com; Matteo Quarantelli matteoquarantelli@maselli.com; Federico Chiesa federicochiesa@maselli.com

Ciao Riccardo, scusa per il ritardo nella mia risposta, ma stavamo aspettando anche delle informazioni da un fornitore. Ti confermo che:

La consegna del materiale sotto-riportato è prevista per il 28/04 UR62 A0840184-CA13AA14ADF6N A0840192-BC11BA11UR29 A0840268-AA14LD21D con modulo ADAM A0117271Nota: dal momento che il deflettore DF15 A0840316-ABD1AA91B è già stato acquistato dal cliente finale, non credo abbiate intenzione di prenderlo, ma per favore confermami

La consegna del materiale sotto-riportato è prevista per il 06/05 Kit Conducibilità A0559015Nota: la settimana prossima potremmo avere un eventuale aggiornamento su un possibile anticipo.

Se per te va bene, ci possiamo sentire alla fine della settimana prossima, per capire quale potrebbe essere lo schedule reale per lattività di prova congiunta presso la nostra sede che hai ipotizzato con Matteo. Fammi sapere. Grazie in anticipo. Approfitto di questa mail per chiederti un aggiornamento sul vostro ordine: sai quando dovrebbe arrivare? Ancora grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Riccardo Riccardo.Guerra@sidel.com

Santini, Paolo Paolo.Santini@sidel.com

Ciao Pietro Mi sai dire, in base ai vostri piani di produzione, quando riuscite ad avere i componenti per poterli provare insieme a voi. Grazie Ciao Riccardo

General

Riccardo Riccardo.Guerra@sidel.com

Santini, Paolo Paolo.Santini@sidel.com

Ok grazie Mi torna adesso anche con la richiesta di Ugo Bini Ciao Riccardo

Bracchi pietrobracchi@maselli.com

Santini, Paolo Paolo.Santini@sidel.com

Ciao Riccardo, anche per lUR62 viene fornito il modulo ADAM; non appare in offerta come riga separata perché linformazione è intrinseca nel modello dellUR62 stesso. Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Riccardo Riccardo.Guerra@sidel.com

Santini, Paolo Paolo.Santini@sidel.com

Grazie Pietro Ho un dubbio soltanto per lUR62 Questo strumento esce direttamente in 4-20mA, giusto? Non ha bisogno del convertitore ADAM come lUR29, lo chiedo perché nello schema condiviso da Matteo ha messo 2 ADAM uno ciascun refrattometro.

Magari mi sono perso qualcosa Grazie Ciao Riccardo

General

Bracchi pietrobracchi@maselli.com

Santini, Paolo Paolo.Santini@sidel.com

Ciao Riccardo, ti confermo che abbiamo preso in carico la tua richiesta di mettere in produzione, prima dellarrivo del vostro ordine ufficiale, il materiale come da offerte che ri-allego, Appena arriverà il vostro ordine, aggiorneremo la nostra conferma dordine (che per ora non vi condividiamo) con il numero dordine stesso e ve la faremo avere. Nei prossimi giorni ti confermeremo la consegna tentativa, in base alle informazioni che ci verranno dalla Produzione. Allego poi la documentazione disponibile per quanto riguarda il Kit Conducibilità. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Riccardo Riccardo.Guerra@sidel.com

Alessandra Bolzoni alessandrabolzoni@maselli.com; Passera, Alessandro Alessandro.Passera@sidel.com; Santini, Paolo Paolo.Santini@sidel.com

Ciao Pietro A seguito della telefonata ti scrivo per chiederti se potete mettere in produzione gli strumenti previsti in questo progetto, Diet As Regular - San Giorgio in Bosco -.

Refrattometro UR62 con convertitore ADAM 4-20mA Refrattometro UR29 con convertitore ADAM 4-20mA Conduttivimetro per misura su acqua Ti chiedo anche un manuale del conduttivimetro per poterlo codificare correttamente ed avere unidea della configurazione che poi quando verremo a testare lo spulceremo bene. Metto in copia anche Paolo per la codifica del convertitore ADAM

Attendo tua conferma il manualetto e il tempo di consegna degli strumenti così pianifichiamo nel dettaglio Grazie Ciao Riccardo General General

Bracchi pietrobracchi@maselli.com

Buongiorno Paolo, in allegato la documentazione richiesta relativa alle due offerte nuovamente allegate, per quanto riguarda lUR62 e lUR29 Essendo materiale dacquisto, non abbiamo invece i disegni relativi al conduttivimetro. Resto a disposizione per ulteriori chiarimenti. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

General

Bracchi

Cantarelli, Paolo Paolo.Cantarelli@sidel.com

Buongiorno Paolo, ho chiesto allufficio Tecnico di verificare quali disegni abbiamo disponibili. Appena mi confermano, la informo. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Bolzoni alessandrabolzoni@maselli.com

Buongiorno Paolo, Le risponderà il collega Pietro Bracchi che sta seguendo il progetto e che ci sta leggendo in copia.

Grazie Cordiali saluti Alessandra

Paolo Paolo.Cantarelli@sidel.com

Buongiorno Alessandra Nel ringraziarla per lofferta inviata, le chiedo cortesemente se potrebbe inviarmi anche la documentazione tecnica ed i disegni / modelli 3D dei componenti sottoelencati affinché possa procedere con la codifica

UR62 e relativi attacchi UR29 e relativi attacchi conduttivimetro e relativi attacchi Grazie Cordiali saluti Paolo Cantarelli Mechanical Engineer - Service Engineering

Sidel s.p.a. Via la Spezia 241/A 43126 Parma, ITALY Tel: +39 0521 999566 www.sidel.com P Please consider the environment before printing this e-mail

General

Bolzoni alessandrabolzoni@maselli.com

Cantarelli, Paolo Paolo.Cantarelli@sidel.com; walter.orsi walter.orsi@teknors.com; Pietro Bracchi pietrobracchi@maselli.com; Matteo Quarantelli matteoquarantelli@maselli.com

di test latenza, offerta RDO250201 per UR29 con modulo Adam e conducibilità

Buongiorno, Come da accordi intercorsi, invio la nostra offerta RDO250201 per:

UR29 solo modulo ADAM (e attacco DN40 a saldare)Kit conducibilità composto da sonda e ricevitore (e nocciolo per installazione). Tempo di consegna: 6 settimane dallordine con possibilità di anticipo in caso di esigenze particolari.

Cordiali saluti

Alessandra Bolzoni

Sales coordinator

Maselli Misure S.p.A. P:

  • 39 0521 257400 E: Parma, Italy

www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Bracchi pietrobracchi@maselli.com

Matteo Quarantelli matteoquarantelli@maselli.com

Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; 'walter.orsi' walter.orsi@teknors.com; 'Passera, Alessandro' Alessandro.Passera@sidel.com; 'Cantarelli, Paolo' Paolo.Cantarelli@sidel.com; Alessandra Bolzoni alessandrabolzoni@maselli.com

Questa volta con lallegato. Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Bracchi

Matteo Quarantelli matteoquarantelli@maselli.com

Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com; Cantarelli, Paolo Paolo.Cantarelli@sidel.com; Alessandra Bolzoni alessandrabolzoni@maselli.com

Ciao Riccardo, per quanto riguarda la prima riga (UR62, vedi parte evidenziata in giallo sotto), puoi fare riferimento allofferta allegata. Come a tua telefonata con Matteo, a breve ti faremo poi avere lofferta per:

UR29 solo modulo ADAM (e attacco DN40 a saldare)Kit conducibilità composto da sonda e ricevitore (e nocciolo per installazione). Grazie, Pietro

Pietro Bracchi Regional Sales Manager

Maselli Misure S.p.A. M:

  • 39 338 5037946 P:
  • 39 0521 257410 E: Parma, Italy www.maselli.com

message with any third party, without a written consent of the sender. If you received this message by mistake, please reply to this message and follow with its deletion, so that we can ensure such a mistake does not occur in the future.

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com; Cantarelli, Paolo Paolo.Cantarelli@sidel.com

Grazie Matteo Ho provato a chiamarti Intendevo dire di tenere i due Adam sui refrattometri, la loro uscita 4-20mA, andrebbe allMP06, lMP06 sarebbe collegato in Profibus ed anche in 4-20mA, verso il PLC, questo per garantire la velocità di trasmissione dai refrattometri, se lMP06 non inserisce latenza nel loop 4-20mA, ed il collegamento Profibus, con lopportuno telegramma vi consentirebbe di ricevere i dati richiesti, ed avremmo un test anche della latenza Profibus su questa macchina. Pietro quando puoi inviarci lofferta aggiornata? Inse4risco nel loop anche Paolo Cantarelli che si occuperà della parte meccanica del progetto, inclusa la codifica dei componenti Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, vedi sotto le risposte. A presto Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Grazie Matteo Bene, lattuatore di che tipo è? Nei 900ms ce anche il suo ritardo di attuazione, anche il tempo di ciclo del PLC ma con una 1200 immagino sia del ms. la sola latenza del loop dellUR potrebbe essere confrontabile con i 300ms attesi. Questo poi lo verifichiamo in campo. OK Direi che sulla prima applicazione verificheremo le prestazioni in campo, ma non mi aspetto problemi anche perché la taglia della macchina non è grande e quindi i flussi sono relativamente bassi. Matteo mi puoi aggiornare lo schema completo? Vedi allegato SLIDE 2 Ha senso inserire comunque lMP06 nel loop di corrente tra i sensori ed il PLC, ingresso 4-20 ed uscita 4-20, si avrebbe un incremento di latenza, non credo sia eccessivo, ma avremmo il vantaggio di poter inviare i dati che avevamo condiviso inizialmente: Intendi al posto dellADAM ? direi di no altrimenti introdurremmo un riardo di alcuni secondi. Se vuoi possiamo predisporre anche lMP06 (SLIDE 1) per una seconda tornata di test. Il test della vita deve essere fatto con gli ADAM (SLIDE 2) ed il log dovrebbe farlo il plc.

centralina MP06 Maselli per gestione Maselli da remoto, alla quale vanno passati alcuni dati: da Sidel a Maselli Alcuni di questi non sono al momento trasferibili da PLC a MP06 (manca telegramma e implementazione sw)

lettura brix UR62 in ingresso linea sciroppo [°Brix] lettura brix UR29 da stream sciroppo [°Brix] temperatura sciroppo [°C] flusso sciroppo [kg/min] flusso acqua [l/min] densità sciroppo [kg/dm3] livello tank sciroppo [%] volume attivo tank sciroppo [L] lettura brix prodotto [°Brix] temperatura prodotto [°C] CO2 prodotto lettura conducibilità dellacqua in ingresso macchina (?) temperatura acqua in ingresso macchina (?) Pressione ingresso linea sciroppo Pressione stream sciroppo

segnali da Maselli a Sidel (Profibus):

Brix sciroppo ingresso linea sciroppo UR62Brix sciroppo stream sciroppo UR29Conducibilità acqua ingressoBrix acqua

Per il resto ti risponde Pietro Chiedo a Pietro se mi può preparare lofferta completa:

UR62 con convertitore RS485/(4-20mA) + attacchi UR29 con convertitore RS485/(4-20mA) + attacchi conduttivimetro con centralina + attacchi vi chiedo inoltre anche i manuali dei sensori, i disegni con le dimensioni, e possibilmente i 3D, tutto questo come sapete per codificare i componenti. Teneteci aggiornati sugli sviluppi del nuovo bus, più veloce, per la comunicazione con gli strumenti. Matteo hai della documentazione oppure è ancora presto? Presto, presto Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, siamo finalmente riusciti a terminare i test con UR29 Di seguito i risultati. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Quarantelli

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Stiamo completando i test .. purtroppo, è stato necessario più tempo del previsto. Spero di avere news domani. Ti tengo aggiornato Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Matteo Dammi buone notizie… Grazie Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ok, grazie. Entro domani dovremmo avere il feedback sui test . Ti faccio sapere Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Matteo Ripeto quello che ho detto andava evidenziato nel momento in cui è emersa la questione, lanno scorso, quando è stato deciso di inserire lopzione profibus ed uscite 4-20mA con la MP06. Certo che non voglio fare verifiche di timing nel progetto di San Giorgio. Ho riassunto uno specchietto per definire il perimetro di cosa stiamo parlando.

Machine size 6 9 10
Prod flow Nom [l/h] 27000 54000 60000
Prod flow Min [l/h] 13500 27000 30000
Prod flow Nom [l/s] 7,50 15,00 16,67
Prod flow min [l/s] 3,75 7,50 8,33
300ms Lt eq @Nom speed 2,25 4,5 5
300ms Lt eq @Min speed 1,125 2,25 2,5

Appena ci vediamo ne parliamo meglio. Dai dati che ho lo strumento in Tolosa UR24, fornisce la stessa misura/valore per almeno 2 secondi, come vedi su macchine molto grandi lindeterminazione sui litri di sciroppo diventano troppi. Fammi sapere per i test Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Ciao Riccardo, purtroppo, con MP06 connesso a 2 sensori è praticamente impossibile soddisfare il requisito di 300 ms nemmeno aumentando la velocità della seriale a 115200 (adesso andiamo a 19200) E un problema intrinseco del protocollo master-slave tra mp06 e sensori. Resta in realtà da verificare se questo requisito, allinterno dellintera catena di pilotaggio sia davvero necessario. Noi potremmo, come anticipato, predisporre anche lMP06 per i test comparativi. Cercheremo di fare i test mercoledì o giovedì Ti faccio sapere Grazie Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com

|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

General

Riccardo Riccardo.Guerra@sidel.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com; walter.orsi walter.orsi@teknors.com; Passera, Alessandro Alessandro.Passera@sidel.com

Grazie Matteo Fammi sapere quando riuscite ad effettuare il test “latenza sensore con convertitore RS485 / 4..20 mA”. Sono sicuro che la tempistica dei 300ms, è emersa sin dai primi incontri in Maselli, è per questa ragione che era stato deciso di fornire lMP06 con le uscite in corrente e Profibus, per poi verificare in cantiere se il Profibus fosse sufficiente pensando alle future applicazioni. Una domanda che mi viene spontanea è possibile aumentare il più possibile il baudrate della seriale? Dato che è una 485 differenziale e la lunghezza del cavo è minima. Gradirei partecipare al test, se possibile. In quelloccasione potremmo vedere nel dettaglio i dati raccolti nel mixer di Tolosa, il primo mixer della nuova generazione sidel, fornito di un UR24 con convertitore Profibus Installato nel seguente modo:

Ricordo, inoltre per la cronaca, che al mio rientro da Tolosa diciamo nel 2012, sono passato in Maselli e parlando di questo argomento con Paolo (il softwarista in “pensione”) mi diceva che questa “scarsa” velocità dellUR24 sarebbe stata risolta, oppure gestibile con la nuova generazione di strumenti equipaggiati con un nuovo microcontrollore. Fatemi sapere, Andrea M. fammi sapere per laltro tema a riguardo degli strumenti da inserire al “volo” per avere dati sui trend del brix, prima del progetto. Ciao Riccardo

General

Quarantelli matteoquarantelli@maselli.com

Andrea Maselli andreamaselli@maselli.com; Andrea Devitis andreadevitis@maselli.com; Pierpaolo Fallini pierpaolofallini@maselli.com

Ciao riccardo ecco le 2 slides. Matteo

Ing. Matteo Quarantelli Technical manager

Maselli Misure S.p.A. P:

  • 39 0521 257450 E: Parma, Italy www.maselli.com